Incubators - Welland range - 80°C
Available with a maximum operating temperature of 80°C and a minimum operating temperature of 10°C above ambient. Both gravity and forced convection models are available.
All units have an integral sealed glass door to facilitate product inspection and are designed for long termy accuracy and reliabiliy. As with all products in the Welland range, a wide choice of control and programming options is available.
Heat up times are excellent and temperature stability, with microprocesor three term control, varies from + 0.2°C to + 0.5°C, depending on the model type.
|
 |
| Model |
WIN 30 |
WIN 60 |
WIN 120 |
WIN 200 |
WIF 30 |
WIF 60 |
WIF 120 |
WIF 200 |
| Maximum temperature (°C) |
80 |
80 |
80 |
80 |
80 |
80 |
80 |
80 |
| Chamber dimensions (mm - h / w / d) |
255
330
320 |
350
392
420 |
450
492
520 |
700
590
520 |
300
292
320 |
400
392
420 |
500
492
520 |
750
792
520 |
| External dimensions (mm - h / w / d) |
470
655
470 |
570
765
570 |
670
865
670 |
920
965
670 |
470
665
470 |
570
765
570 |
670
865
670 |
920
965
670 |
| Chamber capacity (l) |
27 |
58 |
115 |
215 |
28 |
66 |
128 |
230 |
| Weight (kg) |
30 |
45 |
60 |
75 |
30 |
45 |
60 |
75 |
Shelves
Number supplied
Maximum possible
Maximum load / shelf (kg)
Maximum load (kg) |
2
3
10
20 |
2
5
10
30 |
2
9
10
40 |
2
15
10
50 |
2
3
10
20 |
2
5
10
30 |
2
9
10
40 |
2
15
10
50 |
| Power rating at 240V (watts) |
250 |
550 |
675 |
1000 |
250 |
675 |
675 |
1000 |
| Holding power at maximum temperature (watts) |
70 |
95 |
140 |
250 |
115 |
150 |
200 |
300 |
Temperature uniformity
(at maximum temperature as a %) |
+3.5 |
+3.5 |
+3.5 |
+3.5 |
+1.5 |
+1.5 |
+1.5 |
+1.5 |
| Temperature stability with PID control (°C) |
+0.5 |
+0.5 |
+0.5 |
+0.5 |
+0.2 |
+0.2 |
+0.2 |
+0.2 |
Heat up time (minutes) 240V 37°C
60°C
80°C |
6.5
12.5
18 |
6.5
12.5
18 |
6.5
12.5
18 |
8
16
25 |
2.5
8.5
16.5 |
3
7
12 |
3
8.5
15.5 |
4
12
23 |
Recovery time (minutes) with door 37°C
open for 60 seconds 60°C
80°C |
0.5
1
1.5 |
1
2.5
3 |
2
3.5
4.5 |
2.5
4
6 |
1
2
3.5 |
0.5
1
1.5 |
1
1.5
2.5 |
1.5
3
5 |
| Air exchanges (volume - l / hr) |
N/A |
N/A |
N/A |
N/A |
1400 |
1400 |
1400 |
1400 |
| Air exchanges / hour |
N/A |
N/A |
N/A |
N/A |
50 |
21 |
11 |
6 |
Lenton
Parsons Lane, Hope,
Hope Valley, S33 6RB,
UK
Registered in England & Wales - Company No. 01371507
BS EN ISO 9001: 2008 registered company
VAT No: GB592639695
+44 (0) 1433 621515
Fax: +44 (0) 1433 623600