Laboratory Ovens - Welland range - 300°C
A modern range of ovens, with a maximum operating temperature of 300°C, providing a combination of excellent performance and reliability.
Increased power and low thermal mass encased fibre insulation ensure both fast heat up times and reduced recovery times.
Reduced holding power once at set temperature, together with the insulation, makes the range economical. Cool safe outer case temperature.
Both gravity and forced air circulation models are available with a wide choice of control options, allowing the most critical performance criteria to be met. Where processes involve the liberation of flammable vapours, a stoving and curing option is available and for processes involving large amounts of water, a moisture extraction option is available.
|
 |
| Model |
WN 30 |
WN 60 |
WN 120 |
WN 200 |
WF 30 |
WF 60 |
WF 120 |
WF 200 |
| Maximum temperature (°C) |
300 |
300 |
300 |
300 |
300 |
300 |
300 |
300 |
Chamber dimensions
(mm - h / w / d) |
250
330
320 |
350
392
420 |
450
492
520 |
700
590
520 |
300
292
320 |
400
392
420 |
500
492
520 |
750
792
520 |
External dimensions
(mm - h / w / d) |
470
655
470 |
570
765
570 |
670
865
670 |
920
965
670 |
470
665
470 |
570
765
570 |
670
865
670 |
920
965
670 |
| Chamber capacity (l) |
27 |
58 |
115 |
215 |
28 |
66 |
128 |
230 |
| Weight (kg) |
30 |
45 |
60 |
75 |
30 |
45 |
60 |
75 |
Shelves
Number supplied
Maximum possible
Maximum load / shelf (kg)
Maximum load (kg) |
2
3
10
20 |
2
5
10
20 |
2
9
10
40 |
2
15
10
50 |
2
3
10
20 |
2
5
10
30 |
2
9
10
40 |
2
15
10
50 |
| Power rating at 240V (watts) |
750 |
1000 |
1500 |
2250 |
1000 |
1500 |
200 |
2700 |
| Holding power at maximum temperature (watts) |
300 |
480 |
720 |
1160 |
350 |
600 |
800 |
1250 |
Temperature uniformity
(at maximum temperature as a %) |
+2.3 |
+2.3 |
+2.7 |
+3.5 |
+1.0 |
+1.0 |
+1.0 |
+1.0 |
| Temperature stability with PID control (°C) |
+0.5 |
+0.5 |
+0.5 |
+0.5 |
+0.2 |
+0.2 |
+0.2 |
+0.2 |
Heat up time (minutes) 240V to: 37°C
60°C
80°C |
12
26
52 |
12
26
52 |
12
26
52 |
14
29
58 |
4.5
12
25 |
4.5
12
25 |
4.5
12
25 |
5.5
14
30 |
Recovery time (minutes) with door 37°C
open for 60 seconds to: 60°C
80°C |
2.5
5
8.5 |
2.5
5
8.5 |
2.5
5
8.5 |
3
6
10 |
1
2.5
4 |
1
2.5
4 |
1
2.5
4 |
1.5
3
5 |
Air exchanges (volume - l / hr at 100°C)
- with standard fan
- with exhaust fan |
N/A
N/A |
N/A
N/A |
N/A
N/A |
N/A
N/A |
1400
10,000 |
1400
10,000 |
1400
10,000 |
1400
10,000 |
Air exchanges / hour
- with standard fan
- with optional exhaust fan |
N/A
N/A |
N/A
N/A |
N/A
N/A |
N/A
N/A |
50
360 |
21
153 |
11
79 |
6
44 |
Lenton
Parsons Lane, Hope,
Hope Valley, S33 6RB,
UK
Registered in England & Wales - Company No. 01371507
BS EN ISO 9001: 2008 registered company
VAT No: GB592639695
+44 (0) 1433 621515